Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Recall number
Z-0809-2022
Recalling firm
Biomet, Inc.
Product
BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.
Status
Terminated
Initiated
2022-02-14
Posted
not on file
Terminated
2024-09-30
Reason
Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.
Root cause
Under Investigation by firm
Action
A recall notification was sent via courier 24 Feb 2022 to distributors, surgeons, and hospital risk managers. Customers advised to locate and quarantine product, return to Biomet and identify any further customers receiving product. Recalled product is identified using specific lot and item numbers. Additionally, surgeons and hospital risk managers are to follow any specified patient monitoring instructions and return the Certificate of Acknowledgement to Zimmer Biomet. If you have further questions or concerns, please call customer service at 574-371-3071 or send questions via email to CorporateQuality.PostMarket@zimmerbiomet.com
Quantity
702
Distribution
Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands
Lot, serial or model codes
29 lots: Lot Number, UDI Number - [061400, (01)00887868465758(17)310630(10)061400], [061410, (01)00887868465758(17)310723(10)061410], [089370, (01)00887868465758(17)310726(10)089370], [260770, (01)00880304857261(17)310917(10)260770], [503410, (01)00880304857285(17)311011(10)503410], [503420, (01)00880304857261(17)310924(10)503420], [503430, (01)00880304857285(17)311011(10)503430], [503450, (01)00880304857261(17)310924(10)503450], [503460, (01)00880304857285(17)311011(10)503460], [503610, (01)00880304857285(17)310924(10)503610], [539370, (01)00880304857285(17)311105(10)539370], [549910, (01)00887868465758(17)310603(10)549910], [577830, (01)00880304857292(17)310917(10)577830], [577840, (01)00880304857292(17)310922(10)577840], [577860, (01)00880304857292(17)310917(10)577860], [637200, (01)00880304857285(17)311027(10)637200], [796600, (01)00880304857285(17)311105(10)796600], [796610, (01)00880304857261(17)311027(10)796610], [886120, (01)00880304857285(17)310805(10)886120], [886200, (01)00880304857285(17)310805(10)886200], [886210, (01)00880304857285(17)310727(10)886210], [886220, (01)00880304857285(17)310726(10)886220], [886230, (01)00880304857285(17)310813(10)886230], [886240, (01)00880304857285(17)310805(10)886240], [886260, (01)00880304857285(17)310922(10)886260], [886270, (01)00880304857285(17)310922(10)886270], [886290, (01)00880304857285(17)310917(10)886290], [954740, (01)00880304857285(17)310922(10)954740], [954760, (01)00880304857285(17)310922(10)954760]
510(k) numbers
["K141263"]
PMA numbers
not on file
Product code
JDR
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register