Skip to the content

FDA Medical Device Recalls (openFDA)

Remel Inc: Remel Xpect Clostridium difficile Toxin A/B, IVD, 40 clostridium difficile toxin A/B tests, REF 24640. Product Usage: The test kit is a rapid in vitro immunochromatographic test for the direct, qualitative detection of Clostridium difficile Toxin A and/or B in human fecal specimens. The test is intended for use as an aid in diagnosis of Clostridium difficile-associated disease.

Recall number
Z-0807-2013
Recalling firm
Remel Inc
Product
Remel Xpect Clostridium difficile Toxin A/B, IVD, 40 clostridium difficile toxin A/B tests, REF 24640. Product Usage: The test kit is a rapid in vitro immunochromatographic test for the direct, qualitative detection of Clostridium difficile Toxin A and/or B in human fecal specimens. The test is intended for use as an aid in diagnosis of Clostridium difficile-associated disease.
Status
Terminated
Initiated
2012-08-20
Posted
2013-02-12
Terminated
2013-06-04
Reason
The firm is recalling the products due to a potential for false negative test results.
Root cause
Material/Component Contamination
Action
Thermo Fisher sent a recall notification letter with details of the recall on August 20, 2012 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers were requested to follow the instructions provided and fax back the attached "Acknowledgement & Receipt Form" to 1-877-428-1924 (toll free in the US) or 1-913-895-4190 (International). For questions contact the Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).
Quantity
50 units
Distribution
Worldwide Distribution - USA Nationwide and the countries Canada and the United Kingdom
Lot, serial or model codes
Lot 141764
510(k) numbers
not on file
PMA numbers
not on file
Product code
LLH
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register