Skip to the content

FDA Medical Device Recalls (openFDA)

C P Medical: Product is an EO Sterile, Polyester braided and coated nonabsorbable surgical suture with double needle, size 2/0 (3.0 Metric), NSH 26 mm TAPER Needle. Product is wrapped in a five-fold card with inserted flap ends (overwrap). A label is placed on the overwrap and the packet is then inserted into a tyvek/mylar pouch. One end of the overwrap card is printed with 'Open'. This end is positioned in the pouch so it is closest to the peel-open end. No suture is exposed or visible through the overwrap which must be torn to have access to the suture and needle. Product is boxed (12 to a box) and shrink wrapped. Product overwrap label includes: '2/0 (3.0 metric) CP523A Polybond Green Braided, Polyester Coated Nonabsorbable Sterile Suture Rx Only TAPER NSH 36' (90cm) 1/2 26mm Manufactured by CP Medical Portland, OR USA'. Product box label includes: '2/0 (3.0 metric) CP523A NSH 26 mm 1/2 TAPER POLYBOND Green Polyester Braided Coated Nonabsorbable 36' (90cm) STERILE EO Manufacturer CP Medical Portland, OR USA' . Product was distributed to single consignee, a distributor in South Africa.

Recall number
Z-0807-2007
Recalling firm
C P Medical
Product
Product is an EO Sterile, Polyester braided and coated nonabsorbable surgical suture with double needle, size 2/0 (3.0 Metric), NSH 26 mm TAPER Needle. Product is wrapped in a five-fold card with inserted flap ends (overwrap). A label is placed on the overwrap and the packet is then inserted into a tyvek/mylar pouch. One end of the overwrap card is printed with 'Open'. This end is positioned in the pouch so it is closest to the peel-open end. No suture is exposed or visible through the overwrap which must be torn to have access to the suture and needle. Product is boxed (12 to a box) and shrink wrapped. Product overwrap label includes: '2/0 (3.0 metric) CP523A Polybond Green Braided, Polyester Coated Nonabsorbable Sterile Suture Rx Only TAPER NSH 36' (90cm) 1/2 26mm Manufactured by CP Medical Portland, OR USA'. Product box label includes: '2/0 (3.0 metric) CP523A NSH 26 mm 1/2 TAPER POLYBOND Green Polyester Braided Coated Nonabsorbable 36' (90cm) STERILE EO Manufacturer CP Medical Portland, OR USA' . Product was distributed to single consignee, a distributor in South Africa.
Status
Terminated
Initiated
2007-04-03
Posted
2007-05-24
Terminated
2008-08-01
Reason
Potential for sterility of product to be compromised. Suture package seals incomplete or missing.
Root cause
Other
Action
On 04/03/2007 the firm notified its Polyester product (Polybond) customer by telephone and followed up on 04/09/2007 with a letter via certified mail requesting the distributor to discontinue sales of the affected product lot, to notify their customers of the issue, and return stock to CP Medical. The firm provided the distributor with a letter dated 04/09/2007 to be provided to the distributor's customers requested those customers to discontinue use of the affected product lots and return stock to their distributor who in turn are to return product to CP Medical. On 04/20/2007 the firm notified its Polypropylene and Polydioxanone product customers by letter dated 04/17/2007 via certified mail requesting distributors to discontinue sales of the affected product lot, to notify their customers of the issue, and return stock to CP Medical. The firm provided the distributor with a letter dated 04/18/2007 to be provided to the distributor's customers requested those customers to discontinue use of the affected product lots and return stock to their distributor who in turn are to return product to CP Medical.
Quantity
121 boxes (12/box) Lot I0625120, and 121 boxes (12/box) Lot I0627120-01
Distribution
Firm distributes products to human and veterinary medical device distribution centers worldwide. Polybond Polyester sutures, product CP523A, were distributed to one distribution center in South Africa. Polypropelene sutures, product 8665, were distributed to veterinary distribution centers in OH, OR, and NE for nationwide distribution. Polydioxanone sutures, product M398A, were distributed to human and veterinary distribution centers in AZ, IA, OH, LA, MA, MI, MO, NE, and TX for nationwide distribution.
Lot, serial or model codes
Product label is coded CP523A. Lot numbers include: I0625120 expiration 2011-08, and I0627120-01 expiration 2011-08.
510(k) numbers
["K001172"]
PMA numbers
not on file
Product code
GAT
Firm FEI
3002609049
Firm city
Portland
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register