Skip to the content

FDA Medical Device Recalls (openFDA)

Qualigen Inc: The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.

Recall number
Z-0805-2025
Recalling firm
Qualigen Inc
Product
The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.
Status
Open, Classified
Initiated
2024-12-03
Posted
2025-01-03
Terminated
not on file
Reason
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
Root cause
Labeling Change Control
Action
The recalling firm issued letters dated 11/19/2024 via email to their customers explaining the reason for recall and providing the correct expiration dates for each of the affected lot numbers. The risk to health was provided and instructions given to be completed. The instructions were to cease use of the affected product, dispose of all remaining kits from the affected lot numbers, document the number of kits disposed, complete the acknowledgment form by FAX or email within 5 days.
Quantity
426
Distribution
Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
Lot, serial or model codes
UDI-DI: 20816467020345, Lot numbers: 2408018-1 Exp: 2025-09-05; 2409015-1, Exp: 2025-06-27; and 2409015-2, Exp: 2025-06-26.
510(k) numbers
["K182521"]
PMA numbers
not on file
Product code
CDZ
Firm FEI
3003355251
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register