FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Uroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
- Recall number
- Z-0805-2021
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Uroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray diagnosis, endourology and minimal invasive urology/surgery).
- Status
- Terminated
- Initiated
- 2020-09-09
- Posted
- not on file
- Terminated
- 2021-04-16
- Reason
- After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time.
- Root cause
- Device Design
- Action
- On September 9, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that the firm has released an additional safety measure to avoid the potential issue. A safety gear will be installed into the monitor support arm out of an abundance of caution. The corrective action is available and is being provided to affected users free of charge. The firm will contact customers to arrange a date to perform the corrective action. Please feel free to contact the firm to schedule an earlier appointment at 1-800-888-7436.
- Quantity
- 6
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- Serial Nos. 3417 3414 3415 3418 3428 3416
- 510(k) numbers
- ["K101491"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28