Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: GEMINI TF 16 PET/CT System (with software version 3.5.1 and 3.5.1.1), Model # 882470 & GEMINI TF 64 PET/CT System (with software version 3.5.1 and 3.5.1.1), Model #882471 Philips Medical Systems, Highland Heights, OH The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Recall number
Z-0805-2012
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
GEMINI TF 16 PET/CT System (with software version 3.5.1 and 3.5.1.1), Model # 882470 & GEMINI TF 64 PET/CT System (with software version 3.5.1 and 3.5.1.1), Model #882471 Philips Medical Systems, Highland Heights, OH The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Status
Terminated
Initiated
2011-12-02
Posted
2012-01-18
Terminated
2013-02-22
Reason
Philips received notification by a customer of an issue relating to unwanted auto delete occurrences resulting in raw data being deleted and, in some cases, raw data being deleted immediately after scan acquisition but before images are reconstructed.
Root cause
Software design
Action
Philips Medical System sent a Urgent Medical Device Correction letter dated December 5, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The following are recommended actions: 1. Please ask your local Philips Service Engineer to ensure that the Auto Delete function is configured to run automatically. 2. Please share this information with all Philips GEMINI TF system operators at your site. 3. Retain this document with your system's instructions for Use Manual. For further information or support concerning this issue please contact your local Philips representative in the United States and Canada at 1-800-722-9377, and follow the recorded menu options to reach Customer Solutions Engineer; in all other countries please dial your local Philips Healthcare office.
Quantity
197 units
Distribution
Worldwide Distribution -- USA (nationwide) including the states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NY, OH, OR, PA, PR, TN, TX, WA and WV. and the countries of: Argentina, Australia, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Denmark, Egypt, France, Germany, Hungary, India, Israel, Italy, Japan, Netherlands, Pakistan, Philippines, Poland, Russian Fed, Saudi, Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine and United Kingdom.
Lot, serial or model codes
Product Code 882470, Seriel #'s: 7003-7006, 7008-7011, 7014, 7015, 7017-7026, 7031, 7034-7036, 7039, 7041, 7043, 7045, 7047, 7049, 7050-7054, 7058, 7060, 7062, 7064, 7066, 7068, 7069, 7073-7075, 7077, 7080, 7081, 7084-7087, 7089, 7093-7099, 7104, 7106-7108, 7112, 7114, 7115, 7118, 7120-7123, 7125, 7128, 7129, 7131, 7133, 7135, 7137-7140, 7142, 7143, 7145, 7147, 7148, 7156, 7159, 7161, 7163, 7165, 7166, 7168, 7170, 7173, 7174, 7176, 7177, 7181, 7183-187, 7193-7200, 7205, 7208, 7210, 7211, 7215, 7216, 7220, 7223 & 7224.  Product Code 882471, Seriel #'s: 7007, 7012, 7013, 7016, 7027-7030, 7032, 7042, 7044, 7055, 7057, 7059, 7061, 7063, 7065, 7067, 7071, 7072, 7076, 7078, 7079, 7082, 7083, 7088, 7091, 7092, 7102, 7103, 7105, 7109, 7110, 7116, 7117, 7119, 7126, 7132, 7136, 7141, 7144, 7146, 7149, 7152-7154, 7158, 7160, 7164, 7167, 7169, 7171, 7172, 7175, 7178-7180, 7182, 7188-7190, 7192, 7201-7204, 7206, 7209, 7212-7214, 7217-7219 & 7221.
510(k) numbers
["K052640"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register