Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Spine: Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.

Recall number
Z-0803-2015
Recalling firm
Stryker Spine
Product
Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.
Status
Terminated
Initiated
2014-10-06
Posted
2014-12-16
Terminated
2017-05-08
Reason
Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.
Root cause
Nonconforming Material/Component
Action
Stryker Spine sent an Urgent Medical Device Removal Notification/Acknowledgement Form dated 10/6/2014 via Fed Ex. The letter identified the affected product, discussed the product description, product issue, potential hazards, risk mitigations, and actions needed. Customers were asked to read the notice thoroughly and make sure that it is understood. Additionally, they are to complete the enclosed acknowledgement form and fax or e-mail it back.
Quantity
15 units
Distribution
Distributed to GA, IN, MI, NJ, OH, PA, TX & VA.
Lot, serial or model codes
Lot number 4279
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register