Skip to the content

FDA Medical Device Recalls (openFDA)

SALTER LABS: Salter Labs Bubble humidifier, with 6 PSI safety valve, 350cc, Latex Free; Single Patient use; used with supplemental oxygen therapy. Model Numbers 7600-0-50, E7600-0-50, NP7600-0-50 and special order patient kits that contain 7600 humidifiers. Patient kits have brand names of Salter Labs, AdvaCare; Allcare Medical; Memorial Home Services; Therapy Support Inc; Mendo-Lake Home Respiratory Services; MedAssurance Inc.; Abundant Home Care. Product Usage: Humidifier - single patient use. used with supplemental oxygen therapy.

Recall number
Z-0803-2012
Recalling firm
SALTER LABS
Product
Salter Labs Bubble humidifier, with 6 PSI safety valve, 350cc, Latex Free; Single Patient use; used with supplemental oxygen therapy. Model Numbers 7600-0-50, E7600-0-50, NP7600-0-50 and special order patient kits that contain 7600 humidifiers. Patient kits have brand names of Salter Labs, AdvaCare; Allcare Medical; Memorial Home Services; Therapy Support Inc; Mendo-Lake Home Respiratory Services; MedAssurance Inc.; Abundant Home Care. Product Usage: Humidifier - single patient use. used with supplemental oxygen therapy.
Status
Terminated
Initiated
2011-10-21
Posted
2012-01-26
Terminated
2012-02-15
Reason
A manufacturing defect was discovered internally which indicates that in a small number of humidifier lids, the pop-off may not occur at the appropriate pressure and can result in a leak.
Root cause
Process control
Action
Salter Labs contacted each customer directly on October 21, 2011 to inform them of the recall. Upon speaking to the direct customers, Salter Labs also provided a Recall Customer Reply Form for each direct customer to complete and return by fax to 661-854-6816 or 1-800-628-4690. Customers were instructed to immediately examine their inventory of the recalled product, remove and quarantine all affected products, indicate if product would be returned and asked if product was further distributed. If so, the second page of the response form was to be completed and returned. Customers were informed that a representatives from Salter Labs can assist, if needed, in returning all affected products to Salter and completing the attached forms. Salter's direct customers are instructed to contact their customers, should the product have undergone further distribution.
Quantity
6,060 units ( 163 cases)
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AZ, CA, IL, MO, NJ, OH, OK, TX and the country of: Italy
Lot, serial or model codes
Model Numbers 7600-0-50, E7600-0-50, NP7600-0-50 and special order patient kits that contain 7600 humidifiers. Patient kits have brand names of Salter Labs, AdvaCare; Allcare Medical; Memorial Home Services; Therapy Support Inc; Mendo-Lake Home Respiratory Services,; MedAssurance Inc.; Abundant Home Care; Humidifier lot numbers: 091911 and 092611.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BBT
Firm FEI
1314417
Firm city
Arvin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register