FDA Medical Device Recalls (openFDA)

Leonhard Lang Gmbh: GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes. REF 2059144-001 and 2059144-005. . Product Usage: Usage: Skintact Pediatric Multifunction Electrodes are for use on pediatric patients, less than 8 years of age, or weighing less than 25 kg (55 lbs), for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and for single use only.

Recall number
Z-0802-2015
Recalling firm
Leonhard Lang Gmbh
Product
GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes. REF 2059144-001 and 2059144-005. . Product Usage: Usage: Skintact Pediatric Multifunction Electrodes are for use on pediatric patients, less than 8 years of age, or weighing less than 25 kg (55 lbs), for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and for single use only.
Status
Terminated
Initiated
2014-12-03
Posted
2014-12-19
Terminated
2015-08-13
Reason
The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.
Root cause
Component design/selection
Action
The one consignee was sent an Urgent Medical Device Recall letter, dated 12/3/2014. The letter identified affected product, explained the issue, and asked for product to stop being used and distributed, and to return the devices. Questions can be addressed to Bettina Sarlay, Sales Manager, Leonhard Lang, at bettina.sarlay@leonhardlang.at.
Quantity
1080 pieces
Distribution
US Nationwide Distribution in the states of OH and TX.
Lot, serial or model codes
Lots: 30618-0771, 30702-0778, 40217-0776, 40407-0777, 40424-0772, 40704-0775
510(k) numbers
["K081371"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
6736
Firm city
Innsbruck
Firm state
not on file
Firm country
Austria

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register