Skip to the content

FDA Medical Device Recalls (openFDA)

Civco Medical Instruments Inc: Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.

Recall number
Z-0802-2014
Recalling firm
Civco Medical Instruments Inc
Product
Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.
Status
Terminated
Initiated
2013-10-25
Posted
2014-01-21
Terminated
2017-02-02
Reason
Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
Root cause
Device Design
Action
Civco Medical Instruments Inc sent a FIELD SAFETY NOTIFICATION letter dated October 28, 2013 via UPS to their customers. The notification identified the affected device, problem, and field safety corrective action. The letter describes how customers can check to see whether the scales on their device are out of alignment. The letter states CIVCO will rework the system to correct the difference. For questions contact sales representative at CIVCO Medical Solutions, Radiation Oncology at 800-842-8688 or 712-737-8688.
Quantity
138 belly boards
Distribution
Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. Foreign distribution was made to Brazil, Canada, Chile, Czech Republic, France, Germany, Iraq, Israel, Italy, Japan, Malaysia, Saudi Arabia, Slovenia, South Korea, Spain, Turkey, and United Arab Emirates.
Lot, serial or model codes
Lot #'s: M441670, M469970, M469980, M469990, M484910, M484920, M516200, M516210, M516220, M533340, M533350, M533360, M533370, M533380, M533970, M533980, M533990, M534000, M544650, M544660, M544670, M550150, M551490, M554940, M554950, M556540, M556550, M559560, M560740, M575850, M578310, M581730, M591090, M591100, M591110, M591120, M591130, and M599570
510(k) numbers
["K121545"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1937223
Firm city
Kalona
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register