Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.

Recall number
Z-0802-2013
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.
Status
Terminated
Initiated
2010-11-01
Posted
2013-02-12
Terminated
2013-11-08
Reason
Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
Root cause
Software design
Action
Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns.
Quantity
622 units
Distribution
Worldwide Distribution.
Lot, serial or model codes
Catalog number: 101 and Serial Numbers: 0503, 1000, 1017, 1070, 1071, 1073, 1074, 1076, 1077, 1080, 1082, 1083, 1107-1110, 1115-1117,1124,1126-1134, 1136, 1137, 1139-1146, 1148-1152, 1154-1156, 1159-1162, 1164-1166, 1168-1177, 1182, 1185-1190, 1192,1193, 1195, 1197, 1198, 1225, 1228, 1229, 1237, 1239-1241,1243-1245,1247, 1249-1326, 1332-1339, 1344-1347, 1349, 1350-1363, 1365-1409, 1411-1436,1438, 1440-1447, 1449-1453, 1455-1463,1465-1482,1484-1492,1494-1552,1555-1565, 1567-1600,1602,1605-1607,1609-1611,1613-1623,1625-1629,1632-1651,1653, 1655-1660, 1662-1664, 1666, 1667,1669-1679,1681-1684,1686,1689-1691, 1693-1710,1712-1715,1717-1750,1752-1757,1759-1771,1773-1785,1787,1788,1790-1809, 1811-1831,1833,1838,1839,1855-1858,1866, and TK001-TK011.
510(k) numbers
not on file
PMA numbers
not on file
Product code
DRY
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register