Skip to the content

FDA Medical Device Recalls (openFDA)

Guidant Corp-Cpi Division: Guidant EasyTrak 8 French Guiding Catheter, Inner Diameter .087'', 2.21 mm, Overall Length 54 cm, Coronary Sinus - Multi-Purpose Long Tip Shape, CS-MPL, REF 7507

Recall number
Z-0801-04
Recalling firm
Guidant Corp-Cpi Division
Product
Guidant EasyTrak 8 French Guiding Catheter, Inner Diameter .087'', 2.21 mm, Overall Length 54 cm, Coronary Sinus - Multi-Purpose Long Tip Shape, CS-MPL, REF 7507
Status
Terminated
Initiated
2004-03-11
Posted
2004-07-20
Terminated
2005-10-31
Reason
The terminal hub component in a some of the catheters has an unintended abrupt transition between the entry port and the catheter lumen which can result in difficulty introducing or removing delivery system components during lead placement.
Root cause
Other
Action
Starting on March 11, 2003, the firm''s representatives visited consignees, to whom the recalled catheters had been consigned, and retrieved the catheters.
Quantity
20 catheters
Distribution
Nationwide in the United States.
Lot, serial or model codes
Lots 4020551, 4021151, 4021251, and 4022551
510(k) numbers
["K021284","K021455"]
PMA numbers
not on file
Product code
DQR
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register