Siemens Healthcare Diagnostics, Inc.: Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.

Recall number
Z-0800-2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Status
Terminated
Initiated
2013-12-17
Posted
2014-01-21
Terminated
2014-08-26
Reason
An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
Root cause
Software design
Action
The distribution of the Urgent Medical Device Correction letter scheduled to begin on December 17, 2013. x Field service personnel will be notified of the Urgent Medical Device Correction Letter via the Siemens Knowledge Base, as well as information regarding placing labels on the customer system to remind the customer that ///// is not a reportable result. x The Customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue. x Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction.
Quantity
1563
Distribution
Worldwide Distribution. USA nationwide and the country of Canada.
Lot, serial or model codes
ADVIA 1200 with software versions: V2.00 and V2.01  ADVIA 1650 with software versions V3.52 and V4.01
510(k) numbers
["K990346"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register