Skip to the content

FDA Medical Device Recalls (openFDA)

Trilliant Surgical, LLC: Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.

Recall number
Z-0799-2020
Recalling firm
Trilliant Surgical, LLC
Product
Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.
Status
Terminated
Initiated
2019-03-14
Posted
not on file
Terminated
2022-03-16
Reason
A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.
Root cause
Process control
Action
The recall strategy included sending written recall communication and response forms as of 03/14/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device (as the device is considered reusable) as of 03/14/2019. The communication instructed the consignees to: " Check inventory stock for identified devices " Locate and remove and/or discontinue use of identified devices " Return all related product(s) to Recalling Firm in accordance with the included return response / acknowledgement form.
Quantity
20 units
Distribution
US Nationwide distribution in the states of CO, CT, IA, TX and WI. OUS: None
Lot, serial or model codes
Part Number: 320-35-003 Lot Number: TSL006617 UDI: (01)00812926026893(10)TSL006617
510(k) numbers
not on file
PMA numbers
not on file
Product code
HXX
Firm FEI
3007420745
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register