Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.

Recall number
Z-0799-2016
Recalling firm
Cook Inc.
Product
Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
Status
Terminated
Initiated
2016-01-06
Posted
2016-02-24
Terminated
2017-08-23
Reason
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.
Root cause
Process control
Action
On 1/6/2016, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Management/Recall Administration via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
Quantity
1341
Distribution
Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada
Lot, serial or model codes
RPN/Catalog number: C-PMS-250-PED C-PMS-251J-PED C-PMS-300-CHILDRENS-032285 C-PMS-301J-PED C-PMS-301J-PED-BH C-PMS-301-PED C-PMSY-250 C-PMSY-251 C-PMSY-251J C-PMSY-300 C-PMSY-300J C-PMSY-300J-UT-A C-PMSY-300-PED C-PMSY-301J C-PMSY-301J-BH C-PMSY-301J-PED C-PMSY-301J-PED-OTP C-PMSY-400-HURLEY-022388 C-PUM-300J C-PUM-301 C-PUM-301J C-PUMY-301J  Lots: NS6168054 NS6178118 NS6063352 NS5855441 5824486 5898058 5967405 6079119 NS5932737 NS6028554 NS6042799 NS5922707 NS5932811 NS6145002 5987014 NS6178099 NS6028540 NS5830758 NS5866960 NS5918174 NS5974263 NS6042577 NS6047450 NS6090278 NS6105922 NS6124990 NS5881337 NS5892116 NS6031991 5782006 5855241 NS5881336 NS5892113 NS5945282 6062729 NS5970950 NS6108255 NS6153173 5793198 5925744 6010406 6148800 NS5796483 NS5928602 NS6047451 NS5970942 NS5954223 5945120 NS5954224 5973354
510(k) numbers
not on file
PMA numbers
not on file
Product code
DQY
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register