FDA Medical Device Recalls (openFDA)
Teleflex Medical: Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are indicated for airway management.
- Recall number
- Z-0799-2015
- Recalling firm
- Teleflex Medical
- Product
- Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are indicated for airway management.
- Status
- Terminated
- Initiated
- 2014-11-12
- Posted
- 2014-12-12
- Terminated
- 2017-10-13
- Reason
- Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube.
- Root cause
- Process control
- Action
- Consignees were notified by an Urgent Medical Device Recall Notification letter on 11/12/2014. The letter identified the affected product, as well as, the reason for the recall. Customers are instructed to immediately discontinue use of and quarantine any affected product. They are also instructed to complete and fax or e-mail the enclosed Recall Acknowledgement Form regardless of whether they have affected product in stock or not. Upon receipt of the form, a customer service representative will contact customers to provide an RGA number, as well as, instructions for the return of the product. A local sales representative or customer service should be contacted with questions at 1-866-246-6990.
- Quantity
- 14,000 units
- Distribution
- Worldwide Distribution -- U.S., Belgium, Canada, and Japan.
- Lot, serial or model codes
- Product Code: 5-10116, Lot number: 01A1400211
- 510(k) numbers
- ["K822082"]
- PMA numbers
- not on file
- Product code
- BTR
- Firm FEI
- 3005747797
- Firm city
- Durham
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28