Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: Fresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritoneal dialysis.

Recall number
Z-0799-2013
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Fresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritoneal dialysis.
Status
Terminated
Initiated
2012-10-12
Posted
2013-02-11
Terminated
2020-12-30
Reason
There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.
Root cause
Process control
Action
Fresenius Medical Care began contacting the affected customers by telephone and mailing an "URGENT: FMCNA LIBERTY CYCLER SET RECALL" letter and Return Receipt form dated October 31, 2012. The letters identified the specific part number and product lot and request that customers discontinue use of the lots immediately and to return unused product to FMCNA. Instructions are provided for returning the product, asking customers to isolate and discontinue use of the lot number 12BR08009 and place the product in a secure, segregated area. Consignees were instructed to contact their Fresenius Medical Care Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product.
Quantity
560 cases, totaling 5590 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot 12BR08009
510(k) numbers
["K043363"]
PMA numbers
not on file
Product code
FKX
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register