Skip to the content

FDA Medical Device Recalls (openFDA)

Mindray DS USA, Inc. d.b.a. Mindray North America: Mindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, and 300BF-PA00003 (DPM Central Monitoring System) 115-006908-01, -03, -04 (DPM software upgrade kit). The DPM Central Monitoring System transfers information between DPM Central station and other networked devices.

Recall number
Z-0798-2013
Recalling firm
Mindray DS USA, Inc. d.b.a. Mindray North America
Product
Mindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, and 300BF-PA00003 (DPM Central Monitoring System) 115-006908-01, -03, -04 (DPM software upgrade kit). The DPM Central Monitoring System transfers information between DPM Central station and other networked devices.
Status
Terminated
Initiated
2012-11-28
Posted
2013-02-11
Terminated
2013-08-15
Reason
Mindray has identified a software anomaly that may where, in the presence of an asystole, the Telepack ECG waveform may produce a unique step pattern (instead of a flat line) if Lead 1 or V is being utilized for detecting cardiac arrhythmia.
Root cause
Software design
Action
Mindray North America/ Mindray DS USA Inc. sent an "URGENT PRODUCT CORRECTION ACTION" letter dated November 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact your Mindray Service Representative at 1-800-288-2121 for questions relating to this issue.
Quantity
51 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Part No Serial No 300BF-PA00002 EH-17000037 300BF-PA00002 EH-17000032 300BF-PA00002 EH-17000033 300BF-PA00002 EH-25000099 300BF-PA00002 EH-18000043 300BF-PA00003 EH-1A000072 300BF-PA00002 EH-22000088 300BF-PA00003 EH-1A000073 300BF-PA00002 EH-12000060 300BF-PA00002 EH-17000038 300BF-PA00003 EH-21000077 300BF-PA00002 EH-18000058 300BF-PA00002 EH-23000096 300BF-PA00003 EH-19000059 300BF-PA00003 EH-19000060 300BF-PA00002 EH-18000051 300BF-PA00002 EH-12000054 300BF-PA00002 EH-16000029 300BF-PA00002 EH-11000052 300BF-PA00002 EH-17000039 300BF-PA00003 EH-1A000064 300BF-PA00002 EH-18000050 300BF-PA00003 EH-13000078 300BF-PA00003 EH-13000085 300BF-PA00002 EH-18000053 300BF-PA00002 EH-17000036 300BF-PA00002 EH-17000031 300BF-PA00003 EH-17000040 300BF-PA00002 EH-18000054 300BF-PA00002 EH-1A000069 300BF-PA00002 EH-12000062 300BF-PA00002 EH-21000076 115-006908-04 Software Kit 300BF-PA00003 EH-13000083 300BF-PA00003 EH-13000084 300BF-PA00003 EH-22000087 300BF-PA00002 EH-18000056 300BF-PA00002 EH-17000035 300BF-PA00002 EH-18000048 300BF-PA00002 EH-18000052 300BF-PA00003 EH-21000081 300BF-PA00002 EH-17000030 115-006908-03 Software Kit 115-006908-03 Software Kit 115-006908-03 Software Kit 300BF-PA00003 EH-19000062 300BF-PA00002 EH-18000057 300BF-PA00003 EH-22000083 300BF-PA00002 EH-11000051 300BF-PA00002 EH-18000041 300BF-PA00002 EH-21000082.
510(k) numbers
["K080192"]
PMA numbers
not on file
Product code
MSX
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register