Neurovision Medical Products Inc: 2-Channel Laryngeal Surface Electrode (Model LSE500DCS and LSE500DCL) For use as a Surgical stimulator/locator for recurrent laryngeal nerve.

Recall number
Z-0798-2010
Recalling firm
Neurovision Medical Products Inc
Product
2-Channel Laryngeal Surface Electrode (Model LSE500DCS and LSE500DCL) For use as a Surgical stimulator/locator for recurrent laryngeal nerve.
Status
Terminated
Initiated
2010-01-11
Posted
2010-02-22
Terminated
2011-04-21
Reason
This recall has been initiated due to a conductive glue used to attach the lead wire to electrode, which has been found in a small number of electrodes, to cause a short circuit in one of the channels. This is due to migration of the adhesive from one pad to the other creating an electrical short. Use of one of these defective electrodes could lead to a failure to monitor. Our investigation has c
Root cause
Other
Action
URGENT RECALL: Medical Device Letter dated 1/7/10. The letter identified the affected product and provided a list of necessary actions for customers to take. The customers should immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify their customers and notify them at once of this product recall. Their notification to their customers may be enhanced by including a copy of this recall notification letter. This recall should be carried out to the user level. Your assistance is appreciated and necessary to prevent any failure to monitor. Please complete and return the enclosed response form as soon as possible. If you have any questions, call Ryan Rea or Kevin Rea, at 866-815-6999.
Quantity
4143
Distribution
Nationwide: CA CO FL GA IL LA MI ND NY OH PA TX UT WA WI)
Lot, serial or model codes
Lot numbers 113008, 054409, 096709
510(k) numbers
["K003745"]
PMA numbers
not on file
Product code
ETN
Firm FEI
1930870
Firm city
Ventura
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register