Skip to the content

FDA Medical Device Recalls (openFDA)

Orthofix, Inc: Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).

Recall number
Z-0797-2016
Recalling firm
Orthofix, Inc
Product
Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
Status
Terminated
Initiated
2016-01-08
Posted
2016-02-11
Terminated
2016-11-16
Reason
There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
Root cause
Component design/selection
Action
Affected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local Orthofix representative or customer service for further information, replacements, or disposal instructions.
Quantity
36 units
Distribution
Worldwide Distribution -- US, Spain, Italy, and Australia.
Lot, serial or model codes
BS535418105, BS535419118, BS535420030,  BS535419124, BS562325021, BS535418158, BS535418038, BS535418138, BS535418012, BS535419006, BS535420039,  BS535418127,  BS549024010, BS535419141, BS535418081, BS535418111, BS535418117, BS535420041, BS535418144, BS535420081, BS535419151, BS562325006, BS546825008, BS535418002, BS535418013, BS535418131, BS535420006, BS535420010, BS535420026, BS535420047, BS556219012, BS556219026, BS535418109, BS535420082, BS557717009, and  BS562325023.
510(k) numbers
["K151488"]
PMA numbers
not on file
Product code
HXC
Firm FEI
2183449
Firm city
Lewisville
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register