FDA Medical Device Recalls (openFDA)
Orthofix, Inc: Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
- Recall number
- Z-0797-2016
- Recalling firm
- Orthofix, Inc
- Product
- Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
- Status
- Terminated
- Initiated
- 2016-01-08
- Posted
- 2016-02-11
- Terminated
- 2016-11-16
- Reason
- There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
- Root cause
- Component design/selection
- Action
- Affected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local Orthofix representative or customer service for further information, replacements, or disposal instructions.
- Quantity
- 36 units
- Distribution
- Worldwide Distribution -- US, Spain, Italy, and Australia.
- Lot, serial or model codes
- BS535418105, BS535419118, BS535420030, BS535419124, BS562325021, BS535418158, BS535418038, BS535418138, BS535418012, BS535419006, BS535420039, BS535418127, BS549024010, BS535419141, BS535418081, BS535418111, BS535418117, BS535420041, BS535418144, BS535420081, BS535419151, BS562325006, BS546825008, BS535418002, BS535418013, BS535418131, BS535420006, BS535420010, BS535420026, BS535420047, BS556219012, BS556219026, BS535418109, BS535420082, BS557717009, and BS562325023.
- 510(k) numbers
- ["K151488"]
- PMA numbers
- not on file
- Product code
- HXC
- Firm FEI
- 2183449
- Firm city
- Lewisville
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28