FDA Medical Device Recalls (openFDA)
Bard Peripheral Vascular Inc: LUMBAR PUNCTURE TRAY ADULT 22G X3.5 STRL
- Recall number
- Z-0796-2022
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- LUMBAR PUNCTURE TRAY ADULT 22G X3.5 STRL
- Status
- Open, Classified
- Initiated
- 2022-01-20
- Posted
- not on file
- Terminated
- not on file
- Reason
- Due to lumbar puncture tray manometer failure mode of leaking and mating issues.
- Root cause
- Process control
- Action
- Between 01/28 and 02/01/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via email and FedEx 2-day mailings to customers informing them that between 07/02 to 12/15/2021 the firm received 59 customer complaints for lumbar puncture tray manometers with reported failure modes of leaking and mating issues. Customers are asked to take the following actions: 1. Check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facility s destruction process. 2. Share this notification with all users within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If Customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form to the Recalling Firm contact noted on the form confirming acknowledgement of the correction notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process their credit. 5. Indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory was destroyed. 6. Contact their Recalling Firm's representative to assist in this process or they may also contact the listed in the notification for further assistance. For further assistance contact: North American Regional Complaint Center - 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT
- Quantity
- 29,870 devices
- Distribution
- U.S. Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S. (foreign) including states: Costa Rica and Hong Kong.
- Lot, serial or model codes
- Catalog Number: 4306C Lot Numbers/UDI (GTIN, DI + PI): 0001410709 (01)10885403057069(17)220331(10)0001410709 0001411249 (01)10885403057069(17)220331(10)0001411249 0001412995 (01)10885403057069(17)220331(10)0001412995 0001414260 (01)10885403057069(17)220331(10)0001414260 0001414360 (01)10885403057069(17)220430(10)0001414360 0001415852 (01)10885403057069(17)220331(10)0001415852 0001417193 (01)10885403057069(17)220430(10)0001417193 0001415027 (01)10885403057069(17)220430(10)0001415027 0001419336 (01)10885403057069(17)220430(10)0001419336 0001422223 (01)10885403057069(17)220630(10)0001422223 0001425520 (01)10885403057069(17)220630(10)0001425520 0001426674 (01)10885403057069(17)220630(10)0001426674 0001428621 (01)10885403057069(17)220630(10)0001428621 0001429518 (01)10885403057069(17)220630(10)0001429518 0001430118 (01)10885403057069(17)220630(10)0001430118 0001433265 (01)10885403057069(17)220831(10)0001433265 0001433844 (01)10885403057069(17)220930(10)0001433844 0001434817 (01)10885403057069(17)220930(10)0001434817 0001438147 (01)10885403057069(17)221031(10)0001438147
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- PWY
- Firm FEI
- 2020394
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28