FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. This product is sold separately or as part of Kit # 990256. The depth gauge is an instrument in the VariAx Elbow System. Depth gauges are tools used intra-operatively, to aid in correct screw selection. They can be used with a plate, or directly on bone.

Recall number
Z-0796-2010
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. This product is sold separately or as part of Kit # 990256. The depth gauge is an instrument in the VariAx Elbow System. Depth gauges are tools used intra-operatively, to aid in correct screw selection. They can be used with a plate, or directly on bone.
Status
Terminated
Initiated
2009-12-22
Posted
2010-02-19
Terminated
2010-03-16
Reason
The Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings.
Root cause
Process control
Action
Urgent Product Recall letters, dated December 22, 2009, were sent to all Stryker Branches/Agencies. Letters were also sent to hospitals and surgeons that use the VariaAx Elbow System were also notified. All letters were sent by Federal Express. The letters identified the affected product, stated the issue, and the hazards involved. It also asked customers to examine their inventory and hospital locations to identify affected products. All affected products are to be retrieved and returned to their branch or agency warehouse for reconciliation. All products should be reconciled on the attached Product Accountability Form and a copy should be faxed to 201-831-6069 within 5 days of receipt of the notice. Follow the instructions for shipping the affected product back to the firm. Questions should be directed to 201-972-2100.
Quantity
49
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot code: U09984  VariaAx Elbow System Instrument Set - kitted, material transaction lots # U23151, U24021, U25906 and U29985 all contain one piece of the affected recalled catalog number.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTJ
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register