Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic devices that produce cross sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body.

Recall number
Z-0794-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic devices that produce cross sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body.
Status
Terminated
Initiated
2009-05-12
Posted
2010-12-21
Terminated
2010-12-28
Reason
The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.
Root cause
Nonconforming Material/Component
Action
On 05/12/2009, Philips began sending out the URGENT-Field Safety Notice MR Accessory Cart to their consignees. The letter identified the affected product and stated that the drawer rails were made from magnetic material and were not according specification. The hazards involved were also provided along with how to identify the affected product. The firm issued the FCO 78100296 and the firm's Field Service Engineer replaced the magnetic rails with the non-magnetic rails at the consignee's site. The consignees can contact contact the Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and reference FCO 78100296 with any communication.
Quantity
3 units
Distribution
Nationwide Distribution -- IL, MI, and MD.
Lot, serial or model codes
Site numbers: 505048, 505705, and 41445594.
510(k) numbers
["K031815","K041602"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register