Baxter Healthcare Corporation: Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.

Recall number
Z-0793-2020
Recalling firm
Baxter Healthcare Corporation
Product
Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.
Status
Open, Classified
Initiated
2019-12-12
Posted
not on file
Terminated
not on file
Reason
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Root cause
Device Design
Action
The firm disseminated an urgent medical device correction notice on 12/12/2019. The notice explained the issue and provided the following actions be taken: 1. Baxter will be tightening its internal specifications for the D mold to produce Luers that target the mid to upper range of the ISO specification (target date December 2019). Additional actions, as needed, will be determined through investigation. 2. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory. Returning the Baxter customer reply form promptly will prevent you from receiving repeat notices. 3. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 4. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 5. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please distribute this notification to customers and check the associated box on the reply form. For general questions regarding this communication, contact Baxter Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
Quantity
271,014,984 units
Distribution
Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.
Lot, serial or model codes
Product Codes: 2C8519, UDI 00085412048994; 2C8537, UDI 00085412048970; and 2C8541, UDI 00085412071107 All Lots
510(k) numbers
["K961225"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register