Skip to the content

FDA Medical Device Recalls (openFDA)

Richard Wolf Medical Instruments Corp.: Reuseable rotary blade/abrader, product number 899751514. The oval burr frontally hooded is an attachment or accessory to the air-powered device intended for use during surgical procedures to cut hard tissue or bone and soft tissue. General hospital and plastic surgery use.

Recall number
Z-0791-2016
Recalling firm
Richard Wolf Medical Instruments Corp.
Product
Reuseable rotary blade/abrader, product number 899751514. The oval burr frontally hooded is an attachment or accessory to the air-powered device intended for use during surgical procedures to cut hard tissue or bone and soft tissue. General hospital and plastic surgery use.
Status
Terminated
Initiated
2015-11-25
Posted
2016-02-09
Terminated
2017-08-21
Reason
Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the distal or proximal end making the product dysfunctional.
Root cause
Process control
Action
The firm sent an URGENT Medical Device Recall letter dated 11/25/2015 to consignees informing them of the recall and the actions needed to be taken. These included to not use the affected burr and to destroy the device. The user is also instructed to fill out the Recall Action Response Form to acknowledge receipt of the notification. The form should be scanned to Dsantostefano@richardwolfusa.com or mailed to: Dawn Santostefano, Recall Coordinator Richard Wolf Medical instruments 353 Corporate Woods Parkway Vernon Hill, IL 60061
Quantity
2 units
Distribution
Distributed in Washington.
Lot, serial or model codes
Affected lot 4500166254.
510(k) numbers
["K984304"]
PMA numbers
not on file
Product code
HRX
Firm FEI
1418479
Firm city
Vernon Hills
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06