Medacta Usa Inc: Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

Recall number
Z-0791-2013
Recalling firm
Medacta Usa Inc
Product
Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.
Status
Terminated
Initiated
2012-12-17
Posted
2013-02-07
Terminated
2013-02-07
Reason
The tip of the Quadra Trial Broach size 0, broke during a surgery.
Root cause
Device Design
Action
Medacta sent a Urgent Safety Notification letter dated December 13, 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to check inventory for affected product and contact QA@medacta.us.com for an RMA# to return items to Medacta USA. A new version of the recalled product will be provided to all affected customers. For questions call 805-437-7085 x26.
Quantity
26 pieces U.S.A. (226 pieces Worldwide)
Distribution
US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands
Lot, serial or model codes
Lot/Unit Codes: 075023, 075064, 084000, 084001, 084326, 084743, 084815, 085202, 085455
510(k) numbers
["K072857","K082792","K083558","K093944","K103189"]
PMA numbers
not on file
Product code
MEH
Firm FEI
3006639916
Firm city
Camarillo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register