Skip to the content

FDA Medical Device Recalls (openFDA)

Sam Medical Products: SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09

Recall number
Z-0789-2022
Recalling firm
Sam Medical Products
Product
SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09
Status
Terminated
Initiated
2022-01-28
Posted
not on file
Terminated
2023-04-06
Reason
Due to partially assembled valved dressing that has a hole within it instead of the intended non-valved dressing in the package resulting in that the dressing cannot be centered over the open chest wound and create an occlusive seal due to the hole being present.
Root cause
Process control
Action
On 01/27/2022, the firm sent a "MEDICAL DEVICE RECALL" letter via signature required parcel service to distributors and direct internet customers informing them that the chest seal combo package may contain an occlusive dressing that has a hole in the center which would prevent the dressing from performing as an occlusive chest seal. Instructions to: For Distributors - 1. Indicate with a Yes or No response that you have received the SAM Chest Seal Combo Recall Notification (by signature required UPS delivery, email, fax or by phone) and understand the instructions from the Recall Notification. 2. Indicate with a Yes or No response if there have been any complaints or adverse events reported to you regarding the recalled product. If there was a report that a SAM Chest Seal Combo was not able to be used on a patient, please report the details on the Recall Response Form. 3. Affected Product Information - Information on the quantity of recalled devices shipped to your facility and the date the devices were shipped from SAM Medical is included on the Recall Response Form. - Quantity in Inventory Number of recalled SAM Chest Seal Combo products currently in your possession. - Quantity to Return This could be the same as the Quantity in Inventory but could include devices already returned from the sub-distributor or end user. - Quantity Used or Destroyed This is the number of recalled devices that have been used on patients, used in product demonstrations or in training events. - Quantity Further Distributed This is the number of recalled devices that you have sold to Sub-distributors and Direct Customers. 4. Distributor Information Place a mark the in the check box next to each of the statements that are applicable to you as a Distributor. 5. Complete the form by applying your Signature and the date the form was signed, print your name, company affiliation, phone number and email address. Return this Recall Response form to SAM Medical by Fax (503-
Quantity
2,950 packages
Distribution
U.S.: AZ, CA, CO, FL, GA, MA, NY, OR, PA, Puerto Rico, RI, TX, VA, and WA O.U.S.: Canada
Lot, serial or model codes
Model Number: CS203-EN UDI Code: (01)10822045000190(10)Y060321-09(17)270601 (Case Carton) UDI Code: (01)10822045000193(10)Y060321-09(17)270601 (Each Pouch/Package) Lot Number: Y060321-09
510(k) numbers
["K102403"]
PMA numbers
not on file
Product code
NAD
Firm FEI
3023316
Firm city
Tualatin
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register