FDA Medical Device Recalls (openFDA)
Steris Corporation: Vis-U-All High Temperature Tubing 9"x100'. One roll per box; 10 boxes per case 6 rolls/cs Model Number: 881091
- Recall number
- Z-0789-2021
- Recalling firm
- Steris Corporation
- Product
- Vis-U-All High Temperature Tubing 9"x100'. One roll per box; 10 boxes per case 6 rolls/cs Model Number: 881091
- Status
- Terminated
- Initiated
- 2020-12-02
- Posted
- not on file
- Terminated
- 2022-06-21
- Reason
- Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
- Root cause
- Process control
- Action
- Steris issued Urgent Medical Device Recall issued 12/2/20 to Distributors and Customers. Letter states reason for recall, health risk and action to take: 1.Please immediately inspect on-hand inventory for Vis-U-All High Temperature Sterilization Pouches and Tubing. For the full list of affected product and associated lots, please reference Attachment A to this letter. 2. Please complete the Medical Device Recall Response Form included with this Customer Notification Letter and destroy any remaining product in inventory. STERIS will coordinate shipment of replacement product upon receipt of the completed Recall Response Form. Your STERIS Sales Representative can assist you should you have any questions while completing the form. 3. Return the completed Recall Response Form via email to: Regulatory_Compliance@STERIS.com or via fax to (440) 392-8963. If you have questions regarding this matter, please contact Tammy Gentry, Product Manager, at (440) 392-8625 or STERIS Customer Service at 1-800-548-4873.
- Quantity
- 197 cases
- Distribution
- Nationwide Foreign:
- Lot, serial or model codes
- Lot Number: 200408A
- 510(k) numbers
- ["K153540"]
- PMA numbers
- not on file
- Product code
- FRG
- Firm FEI
- 1527821
- Firm city
- Mentor
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as