Skip to the content

FDA Medical Device Recalls (openFDA)

Linvatec Corp. dba ConMed Linvatec: ConMed Linvatec Biomaterials, PopLok bone Punch PKL-35M, 3.5mm Orthopedic manual surgical instrument - Bone Punch

Recall number
Z-0789-2013
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
ConMed Linvatec Biomaterials, PopLok bone Punch PKL-35M, 3.5mm Orthopedic manual surgical instrument - Bone Punch
Status
Terminated
Initiated
2012-11-15
Posted
2013-02-05
Terminated
2013-05-22
Reason
ConMed Linvatech recalled one lot of their PopLok bone Punch 3.5mm because the metal shaft diameter is 4.5mm when it should be 3.5mm.
Root cause
Employee error
Action
ConMed Linvatech issued a "URGENT MEDICAL DEVICE RECALL NOTIFICATION" to its customers on November 15, 2012 for a recall that was conducted at the end user/distributor level. All recall notifications were sent FedEx next day delivery or equivalent method to notify all direct consignees. The notifications identified the affected product along with reason for the recall health risk. All accounts were asked to immediately check their facility for the affected product, stop using any of the product with specified lot number, and to segregate the product. The letter also asked customers to return all affected product to ConMed Linvatec and respond using a Reply Form. If customers have any additional questions, they should contact the Customer Service department at 800-237-0169. For questions regarding this recall call 727-399-5205.
Quantity
8
Distribution
Worldwide Distribution - USA including California and Internationally to Spain and Korea
Lot, serial or model codes
Lot 256058
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register