Skip to the content

FDA Medical Device Recalls (openFDA)

Avanos Medical, Inc.: Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.

Recall number
Z-0788-2020
Recalling firm
Avanos Medical, Inc.
Product
Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.
Status
Terminated
Initiated
2019-11-05
Posted
not on file
Terminated
2023-08-16
Reason
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.
Root cause
Manufacturing material removal
Action
Consignees were sent an Urgent Product Advisory Notice letter beginning 11/05/2019 via email and overnight mail. The letter identified affected catalog codes, stated reason for recall, stated that blockage is detectable before catheter use by assessing suctioning effectiveness, and asked for a response form to be returned within 5 business days. questions in the US can be directed to avanos8446@stericycle.com. Questions from outside the US can be sent to avanos8446OUS@stericycle.com.
Quantity
114,300 units (37,880 US, 76,420 OUS)
Distribution
Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
Lot, serial or model codes
REF 195 Lot Code / Expiration Date:  M18115T401/2023-APR, M18121T402/2023-MAY, M18128T404/2023-MAY, M18143T401/2023-MAY, M18149T403/2023-MAY, M18156T402/2023-JUN, M18170T402/2023-JUN, M18177T402/2023-JUN, M18192T402/2023-JUL, M18206T401/2023-JUL, M18220T402/2023-AUG, M18247T402/2023-SEP, M18254T405/2023-SEP, M18282T404/2023-OCT, M18311T401/2023-NOV, M18318T402/2023-NOV, M18345T403/2023-DEC, M1834T403/2023-DEC, M18358T403/2023-DEC, M19008T401/2024-JAN, M19058T402/2024-FEB, M19078T403/2024-MAR, M19086T401/2024-MAR, M19098T403/2024-APR, M19134T404/2024-MAY,   REF 201 Lot Code / Expiration Date:  M18044T405/2023-FEB, M18311T402/2023-NOV, M18358T404/2023-DEC, M19037T402/2024-FEB, M19065T402/2024-MAR, M7172T402/2022-JUN.   REF 202 Lot Code / Expiration Date:  M18143T403/2023-MAY, M18192T401/2023-JUL, M18254T404/2023-SEP, M18274T401/2023-OCT, M18296T405/2023-OCT, M18318T401/2023-NOV, M19072T401/2024-MAR, M19142T401/2024-MAY, M7346T403/2022-DEC.   REF 195-5 Lot Code / Expiration Date:  20000095/2024 - JUL, M18274T402/2023-OCT, M18290T402/2023-OCT, M18296T406/2023-OCT, M18311T403/2023-NOV, M18324T401/2023-NOV, M18324T402/2023-NOV, M18332T401/2023-NOV, M18339T401/2023-DEC, M18345T404/2023-DEC, M18352T402/2023-DEC, M19024T402/2024-JAN, M19030T401/2024-JAN, M19051T401/2024-FEB, M19059T401/2024-FEB, M19065T401/2024-MAR, M19072T402/2024-MAR, M19079T401/2024-MAR, M19085T402/2024-MAR, M19098T404/2024-APR, M19099T401/2024-APR, M19105T403/2024-APR, M19120T403/2024-APR, M19128T401/2024-MAY, M19128T403/2024-MAY, M19135T401/2024-MAY, M19142T402/2024-MAY, M19149T405/2024-MAY, M19162T403/2024-JUN, M19177T402/2024-MAY, M19184T401/2024-JUL.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BSY
Firm FEI
3011270181
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register