Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Orthopaedics, Inc.: SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.

Recall number
Z-0787-2016
Recalling firm
DePuy Orthopaedics, Inc.
Product
SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.
Status
Terminated
Initiated
2016-01-08
Posted
2016-02-09
Terminated
2017-03-24
Reason
Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space. Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.
Root cause
Component design/selection
Action
On 1/8/2016, URGENT INFORMATION RECALL NOTICE for Specific Lots of SIGMA High Performance (HP) MBT Non-Keel Punch Knee Instrument notifications were sent to the affected distributors and Medical Professionals with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc.s Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. 5 p.m. EDT). For questions about the device recall information provided, please contact Kim Earle, Senior Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. 5 p.m. EDT).
Quantity
351
Distribution
Worldwide Distribution-US Nationwide, Armenia, Australia, Austria, Belgium, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, P R China, Portugal, Poland, Russia, Singapore, Slovenia, South Africa, Spain, Sweden, South Korea, Switzerland, Thailand, Turkey, UAE, UK, and Brazil.
Lot, serial or model codes
Cat. No. GTIN Lot Description  950502016 10603295227366 ABB92402 SIGMA HP MBT Cemented Punch size 1-1.5  950502017 10603295227373 ABB84491 SIGMA HP MBT Cemented Punch size SZ 2-3  950502018 10603295227380 ABB92358 SIGMA HP MBT Cemented Punch size SZ 4-7  950502020 10603295227403 ABC33125 SIGMA HP MBT Noncemented Punch size 2-3  950502020 10603295227403 ABB81733 SIGMA HP MBT Noncemented Punch size SZ 2-3  950502020 10603295227403 ABB81732 SIGMA HP MBT Noncemented Punch size SZ 2-3  950502021 10603295227410 ABB39952 SIGMA HP MBT Noncemented Punch size SZ 4-7  950502021 10603295227410 ABC33123 SIGMA HP MBT Noncemented Punch size SZ 4-7
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRY
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register