Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Incorporated: TomoTherapy Treatment Delivery System with iDMS - Product Usage: used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

Recall number
Z-0785-2021
Recalling firm
Accuray Incorporated
Product
TomoTherapy Treatment Delivery System with iDMS - Product Usage: used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.
Status
Terminated
Initiated
2020-12-10
Posted
not on file
Terminated
2022-09-26
Reason
"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.
Root cause
Software Design Change
Action
Urgent Medical Device Recall "MLC tickle error" issue during TomoHelical deliver was sent to customers beginning 12/10/2020. Accuray is planning software releases that will address this issue. All TomoTherapy systems at affected software versions will be upgraded to these versions once released.
Quantity
8 units
Distribution
Worlwide distribution - US Nationwide distribution including in the states of CA, FL, IL, ND, NJ, OK, UT and the countries of Australia, China, Italy, Switzerland, United Arab Emirates.
Lot, serial or model codes
Product Codes: 104430, 1018286, H-0000-0003, 1018283. System Version: 1.2 (also referred to as 5.6.2). Serial Numbers: 110105, 110219, 110221, 110267, 110284, 110290, 110308, 110368,
510(k) numbers
["K171837"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register