FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
- Recall number
- Z-0785-2019
- Recalling firm
- Medtronic Perfusion Systems
- Product
- RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
- Status
- Open, Classified
- Initiated
- 2018-11-27
- Posted
- not on file
- Terminated
- not on file
- Reason
- Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
- Root cause
- Process control
- Action
- The firm, Medtronic, began notifying US consignees on 27-Nov-2018 via hand delivery to be followed up with a UPS 2-Day communication titled "URGENT MEDICAL DEVICE RECALL" to each listed consignee that does not confirm hand delivery. Communications OUS also began 27-Nov-2018 using locally approved methods. The letter described the product, problem and actions to be taken. Customers were requested to do the following: 1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary. 3. Complete the enclosed Customer Confirmation Certificate and email to RS.CFQFCA@medtronic.com. If you have any questions about this action, please contact your Medtronic representative or call Director, Cardiac and Vascular Group, at 763-526-2513 or email: carlos.j.alfonzo@medtronic.com.
- Quantity
- 1100 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.
- Lot, serial or model codes
- BT920: GTIN/UDI Lot/Serial Numbers 10673978522460 216044420 10673978522460 216062719 10673978522460 216438180 BT926: GTIN/UDI Lot/Serial Numbers 10673978522477 215782209 10673978522477 215903433 10673978522477 216034158 10673978522477 216107790 10673978522477 216372963 10673978522477 216438301
- 510(k) numbers
- ["K803277"]
- PMA numbers
- not on file
- Product code
- CAC
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28