Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Bard Distaflo Vascular Bypass Graft; Product Code: DF6006SC; 8 mm x 60 cm. Intended for bypass or reconstruction of peripheral arterial blood vessels.

Recall number
Z-0785-2016
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Distaflo Vascular Bypass Graft; Product Code: DF6006SC; 8 mm x 60 cm. Intended for bypass or reconstruction of peripheral arterial blood vessels.
Status
Terminated
Initiated
2016-01-05
Posted
2016-02-09
Terminated
2016-03-25
Reason
Size mislabeling; the graft is labeled as a 6 mm diameter graft when the actual packaged product is an 8 mm graft.
Root cause
Error in labeling
Action
Customer notification letters were sent on 01/05/2016. Customers were asked to not use the affected product, or further distribute it. Customers were asked to check all inventory locations for the affected product code, lot number combination listed. If customers further distributed any of the product they are requested to contact that location, advise them of the recall, and forward the listed instructions. The letter states to remove all identified product from the shelves. If no affected product is going to be returned the firm states to complete and return the attached form stating so. If customers have product it is requested that they also complete the attached form, and state the quantities and lot numbers in stock. The letter requests that customers call the firm's Recall Coordinator, Raye Seisinger, at 1-800-321-4254, option #2 ext. 2501 (M-F 6am to 3pm MST) or email raye.seisinger@crbard.com. Once all information has been verified the customer will be issued a Return Authorization or Consignment Recall Number to facilitate the expedient return of the product. The firm will provide replace product for returned product. Fax the attached form to the firm at 1-800-994-6772. If the customer cannot fax the firm they request that the Recall Coordinator be contact and provided the information verbally. The letter states a mailing label is enclosed to return the affected product. Customers are requested to amek the package as "recalled product" and include the XC or XH number. All products should be returned to the following shipping address: Bard Peripheral Vascular, Inc. 1415 W. 3rd St. Tempe, AZ 85281
Quantity
4 units
Distribution
Distributed in the states of AL, TX, and CA.
Lot, serial or model codes
Product Code: DF6006SC Lot Number: VTZE1782
510(k) numbers
["K983861"]
PMA numbers
not on file
Product code
DSY
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register