Skip to the content

FDA Medical Device Recalls (openFDA)

Datex - Ohmeda, Inc: GE Healthcare Engstrom Carestation Critical Care Ventilator with the High Performance Display Unit with SW version 3.x and 4.11; identification numbers 1505-9000-000 and M1057699 (upgrade). Datex-Ohmeda, Inc., a General Electric Company, doing business as GE Healthcare, P.O. Box 7750, Madison, WI 53707-7550, USA, www.gehealthcare.com.

Recall number
Z-0784-2007
Recalling firm
Datex - Ohmeda, Inc
Product
GE Healthcare Engstrom Carestation Critical Care Ventilator with the High Performance Display Unit with SW version 3.x and 4.11; identification numbers 1505-9000-000 and M1057699 (upgrade). Datex-Ohmeda, Inc., a General Electric Company, doing business as GE Healthcare, P.O. Box 7750, Madison, WI 53707-7550, USA, www.gehealthcare.com.
Status
Terminated
Initiated
2007-02-02
Posted
2007-05-10
Terminated
2008-10-16
Reason
There are two unique conditions with the Engstrom Carestation that require corrective action. These include an anomaly in the control of the safety relief valve and a situation where the display unit may reset.
Root cause
Other
Action
Device Correction Letter to customers was mailed on 2/2/07 giving information about the correction and a short term recommendation until version 4.15 could be installed by a GE representative who would be contacting the customer.
Quantity
446
Distribution
GA, RI, WI, TX. Foreign - AUSTRALIA, BRUNEI DARUSSALAM, BRAZIL, COLOMBIA, COSTA RICA, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDIA, ITALY, KOREA, MEXICO, MALAYSIA, PHILIPPINES, SAUDI ARABIA, SWEDEN, SINGAPORE, GERMANY.
Lot, serial or model codes
Serial numbers: CBCH00101, CBCH00110, CBCJ00135, CBCJ00136, CBCJ00207, CBCJ00472, CBCJ00623, CBCJ00685, CBCJ00686, CBCJ00688 thruCBCJ00702, CBCK00100, CBCK00101 thru CBCK00208, CBCK00210 thru CBCK00269, CBCK00271 thru CBCK00298, CBCK00301 thru CBCK00516, CBCK00519 thru CBCK00526, CBCK00535
510(k) numbers
["K060862"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2183066
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register