Diasorin Inc.: Trep-Chek Syphilis Test Kit, 960 tests, Catalog No. TP-960, Distributed by DiaSorin, Inc. 1951 Northwestern Ave, Stillwater, MN 55082-0285 USA. Manufactured by Phoenix Bio-Tech Corp., 6910 Kitimat Road, Unit #1, Mississauga, Ontario, Canada L5N 5M2.

Recall number
Z-0784-05
Recalling firm
Diasorin Inc.
Product
Trep-Chek Syphilis Test Kit, 960 tests, Catalog No. TP-960, Distributed by DiaSorin, Inc. 1951 Northwestern Ave, Stillwater, MN 55082-0285 USA. Manufactured by Phoenix Bio-Tech Corp., 6910 Kitimat Road, Unit #1, Mississauga, Ontario, Canada L5N 5M2.
Status
Terminated
Initiated
2005-03-16
Posted
2005-04-28
Terminated
2006-07-22
Reason
Package insert incorrectly states that specimen diluent is ready to use, however specimen diluent bottle in kit states that the diluent must be diluted to one liter.
Root cause
Other
Action
Customers were contacted via telephone, informed of the incorrect package insert, asked to destroy any incorrect inserts and a correct insert was faxed to customers.
Quantity
14 kits
Distribution
MA, CA, TX, KS, NY, FL
Lot, serial or model codes
lot # 041207T expiration date 9-30-2005
510(k) numbers
["K001552"]
PMA numbers
not on file
Product code
LIP
Firm FEI
2182595
Firm city
Stillwater
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register