FDA Medical Device Recalls (openFDA)
The Anspach Effort, Inc.: Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.
- Recall number
- Z-0783-2013
- Recalling firm
- The Anspach Effort, Inc.
- Product
- Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.
- Status
- Terminated
- Initiated
- 2012-09-05
- Posted
- 2013-02-04
- Terminated
- 2014-08-13
- Reason
- Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Anspach Effort, Inc. sent an "URGENT: Medical Device Removal" letter dated November 7, 2012, to its customer. The letter identified the product, problem and actions to be taken. The customer was instructed to do the following: 1) Screen their inventory and remove and return all products immediately. 2) Complete and return the attached Customer Reply Form via fax or email to the address provided on the form. Should the customer have any questions, please contact Anspach Product Support at (800) 327-6887.
- Quantity
- 8 devices
- Distribution
- Distributed in the state of Massachusetts
- Lot, serial or model codes
- Lot number: 2000930
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HTT
- Firm FEI
- 1045834
- Firm city
- Palm Beach Gardens
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28