Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: HiSpeed FX/I (2200997-2 with Gantry mo. #2249696)

Recall number
Z-0781-2011
Recalling firm
GE Healthcare, LLC
Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2249696)
Status
Terminated
Initiated
2010-02-01
Posted
2011-01-04
Terminated
2014-05-21
Reason
A defect with the hardware and software displaying the incorrect dose data.
Root cause
Radiation Control for Health and Safety Act
Action
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Quantity
102 total
Distribution
Nationwide Distribution
Lot, serial or model codes
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register