Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker GmbH: Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050

Recall number
Z-0780-2020
Recalling firm
Stryker GmbH
Product
Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050
Status
Terminated
Initiated
2019-12-05
Posted
not on file
Terminated
2020-10-09
Reason
Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.
Root cause
Under Investigation by firm
Action
Stryker issued Urgent Medical Device Recall Letter (PFA# 2219868) to US consignees on 5-Dec-2019 , letter states reason for recall, health risk and action to take: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply. An OUS customer letter was also issued at the country level.
Quantity
433 (289 units U.S. / 144 units OUS)
Distribution
Worldwide distribution - US Nationwide distribution and countries of Switzerland, France, Spain, United Kingdom, Italy, Sweden, Netherlands, Australia.
Lot, serial or model codes
All lots UDI: (01)07613327350210
510(k) numbers
not on file
PMA numbers
not on file
Product code
HRS
Firm FEI
3002807830
Firm city
Selzach
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register