Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Recall number
Z-0779-2023
Recalling firm
Steris Corporation
Product
VERIFY Dual Species Self-Contained Biological Indicators, 100 per box Item Number: S3061 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes
Status
Open, Classified
Initiated
2022-12-12
Posted
2023-01-03
Terminated
not on file
Reason
Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.
Root cause
Employee error
Action
Steris issued Urgent Medical Device Recall letters to Customers and Distributors on 12/12/2022 via FedEX. Letter states reason for recall, health risk and action to take: 1. Please immediately inspect your on-hand inventory for product affected by this recall. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. STERIS will coordinate shipment of replacement product upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: Regulatory_Compliance@STERIS.com or via fax to (440) 392-8963. If you have questions regarding this recall, please contact Melissa Gonsalves, Senior Product Manager at (440) 392-7043, or STERIS Customer Service at 1-800-548-4873.
Quantity
857 boxes
Distribution
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA
Lot, serial or model codes
UDI: 10724995023833, 50724995023824 lot #230613
510(k) numbers
["K083665"]
PMA numbers
not on file
Product code
FRC
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register