Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Configuration Transfer Cable; an accessory for the Colleague Infusion pump, Product code 2M8155; Baxter Healthcare Corporation, Deerfield, IL

Recall number
Z-0778-05
Recalling firm
Baxter Healthcare Corp.
Product
Configuration Transfer Cable; an accessory for the Colleague Infusion pump, Product code 2M8155; Baxter Healthcare Corporation, Deerfield, IL
Status
Terminated
Initiated
2005-03-02
Posted
2005-04-27
Terminated
2006-05-23
Reason
The Configuration Tranfer Cable for the Colleague Infusion Pumps has incorrect thumb screws on it., not allowing firm connections of the cable to the pump to transfer data to the biomedical engineer's diagnostic device.
Root cause
Other
Action
Recall letters dated 3/2/05 were sent to all customers receiving Configuration Transfer Cables between June and November 2004. The letters were sent to the attention of the Biomedical Engineer, and informed them that the thumbscrews were not manufactured according to Baxter''s specifications and were too narrow to secure the cable to the pump. The accounts were requested to return their Configuration Transfer Cables for replacement.
Quantity
404 cables
Distribution
Nationwide, and internationally to Brazil, Canada, Chile, China, Colombia, Hong Kong, Korea, New Zealand, Japan, Taiwan, Saudi Arabia and Turkey.
Lot, serial or model codes
Product code 2M8155, batch #50604
510(k) numbers
not on file
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register