Beckman Coulter, Inc.: BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.
- Recall number
- Z-0777-2023
- Recalling firm
- Beckman Coulter, Inc.
- Product
- BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.
- Status
- Open, Classified
- Initiated
- 2022-11-18
- Posted
- 2023-01-03
- Terminated
- not on file
- Reason
- Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN 800-7212
- Root cause
- Under Investigation by firm
- Action
- The firm issued a MEDICAL DEVICE RECALL notice to its consignees on 11/18/2022 by mail and email. The notice explained the problem and requested the following actions be taken: "If IRISpec CB controls pass (QC results are within QC acceptance criteria): Continue to process patient samples per IFU (P/N 301-7146BE) If IRISpec CB controls fail (QC results are outside QC acceptance criteria): Refer to Addendum A below for troubleshooting No retrospective review of QC or patient sample results is necessary." "Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter."
- Quantity
- 3657 units
- Distribution
- US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan
- Lot, serial or model codes
- UDI/DI 10837461002611, Lot 565-21: 11-30-2022
- 510(k) numbers
- ["K072640"]
- PMA numbers
- not on file
- Product code
- JJW
- Firm FEI
- 3000203293
- Firm city
- Miami
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28