Baxter Healthcare Corporation: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), Product Code 35700BAX
- Recall number
- Z-0775-2025
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), Product Code 35700BAX
- Status
- Open, Classified
- Initiated
- 2024-11-19
- Posted
- 2024-12-27
- Terminated
- not on file
- Reason
- Devices were identified as released after repair without full testing being performed, which includes flow testing.
- Root cause
- Process control
- Action
- Customers were contacted via telephone beginning 11/19/24. Baxter Healthcare is calling to notify you of a potential issue with the Spectrum infusion pump. One of your devices has been identified as released after repair without full testing being performed, which includes flow testing. Baxter is requesting the return of the impacted infusion pump to perform all tests required and confirm that the infusion pump performs as expected. As part of this process, we need to make you aware of the potential hazard associated with flow related issues. Flow issues may cause excessive therapy, free-flow condition, insufficient therapy or interruption of therapy. These may lead to serious adverse health consequences, including permanent injury and death. Send signed form to Global CORP Service Quality Field Actions Global_CORP_Service_Quality_Field_Actions@baxter.com
- Quantity
- 1 units
- Distribution
- US Nationwide distribution in the states of AL, IN.
- Lot, serial or model codes
- GTIN 00085412091570; Serial Number: 779515
- 510(k) numbers
- ["K133801"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28