Skip to the content

FDA Medical Device Recalls (openFDA)

Lab Vision Corporation: Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation, Fremont, CA Thermo Fisher Scientific Anatomical Pathology Tudor Rd., Manor Park Rancorn, Cheshire WA7 1TA UK Analyte specific reagent. It is intended for qualitative immunohistochemistry with normal and neoplastic formalin-fixed, paraffin embedded tissue sections, to be viewed by light microscopy.

Recall number
Z-0775-2014
Recalling firm
Lab Vision Corporation
Product
Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision Corporation, Fremont, CA Thermo Fisher Scientific Anatomical Pathology Tudor Rd., Manor Park Rancorn, Cheshire WA7 1TA UK Analyte specific reagent. It is intended for qualitative immunohistochemistry with normal and neoplastic formalin-fixed, paraffin embedded tissue sections, to be viewed by light microscopy.
Status
Terminated
Initiated
2013-12-06
Posted
2014-01-16
Terminated
2014-02-25
Reason
Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues.
Root cause
Under Investigation by firm
Action
Lab Vision Corporation part of Thermo Fisher Scientific sent an Urgent Medical Device Recall letter dated December 6, 2013, to all affected customers. Revised letters will be sent on January 8, 2014. Changes include that the reagent is an Analyte Specific Reagent (ASR) and not an in vitro diagnostic device in the US. They also clarify that a false positive was reported by a customer and a false negative was confirmed during in house testing. Letters also correct the expiry date from 12/2013 to 03/2015. Data sheet was also updated to eliminate information stating that the presence of ER in breast tumors indicates an increased likelihood of response to estrogen therapy and also that the positive control must be determined by the investigator. If you have any questions, please contact Technical Service at 1-800-522-7270, option 2, option 7.
Quantity
5
Distribution
Worldwide Distribution: US (nationwide) including the states of TX and MA and internationally to: Turkey.
Lot, serial or model codes
PN/ MS-354-P0, LN 354P1206K; Expiration 03/2015.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MVU
Firm FEI
3000719701
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register