Boston Scientific Corporation: AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x15mm UPN: M00553650 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.

Recall number
Z-0774-2021
Recalling firm
Boston Scientific Corporation
Product
AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x15mm UPN: M00553650 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Status
Terminated
Initiated
2020-12-01
Posted
not on file
Terminated
2023-07-14
Reason
The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity
Root cause
Process control
Action
Boston Scientific issued customer notification letters to consignees via Federal Express Priority Overnight mail on December 1, 2020. The letter states reason for recall, health risk and action to take: Cease further distribution or use of any remaining product affected by this removal immediately. Verify by product lot/batch number in the product table whether any product within your inventory is affected. If so, indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions. If you are a distributor, please note that the depth of this removal is to the end user level and this notification should be forwarded to your customers. If you are a facility that has sent products to another hospital or facility within your network, please ensure that this notification is forwarded to them. Your local Sales Representative can answer any questions that you may have regarding this notification.
Quantity
32 units
Distribution
US Nationwide distribution - Foreign: TBD.
Lot, serial or model codes
Lot Number: 26152465, 26152466, 26152467  Exp. Date: 10/8/2022 GTIN: 08714729904595
510(k) numbers
["K181905"]
PMA numbers
not on file
Product code
PCU
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register