Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Dimension(R) Creatinine Flex(R) Reagent Cartridges. The CREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista System.

Recall number
Z-0772-2012
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension(R) Creatinine Flex(R) Reagent Cartridges. The CREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista System.
Status
Terminated
Initiated
2011-11-22
Posted
2012-01-15
Terminated
2014-04-03
Reason
Siemens has confirmed the potential for falsely elevated creatinine results with CREA FLex(R) reagent cartridge lots GA2196, GA2203, DC2221, DB2221, GA2229 and FA2237 when processing tests from a well that has been punctured.
Root cause
Under Investigation by firm
Action
Siemens sent an "URGENT FIELD SAFETY NOTICE" dated November 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter recommends that customers discontinue and discard any remaining inventory of the affected product. The firm also recommends customers to test QC at least every 8 hours with lots manufactured after lot FA2237. Additionally, a Field Correction Effectiveness Check form was attached to the letter for customers to complete and return via fax to 302-631-8467. Contact the Siemens Customer Service Center at 888-588-3916 for questions regarding this notice.
Quantity
not on file
Distribution
Worldwide Distribution-USA (nationwide) including the District of Columbia and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Benin, Bosnia Herzegovina, Botswana, Brazil, Canada, Canary Islands, Ceuta/Mililla, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, France, French Guinea, Germany, Great Britian, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Mexico, Montenegro, Morocco, Nepal, Nether Antilles, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, Yemen and Zimbabwe.
Lot, serial or model codes
Catalog number DF33A and lot numbers GA2196, GA2203, DC2221, DB2221, GA2229 and FA2237.
510(k) numbers
["K925668"]
PMA numbers
not on file
Product code
CGX
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register