Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile, Intermittent self-catheterization

Recall number
Z-0771-2017
Recalling firm
Teleflex Medical
Product
RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile, Intermittent self-catheterization
Status
Terminated
Initiated
2016-11-22
Posted
not on file
Terminated
2017-11-24
Reason
Labeling Inconsistency: The products have been labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.
Root cause
Error in labeling
Action
Teleflex sent an Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers with affected stock, immediately discontinue use and quarantine any affected products. To return product, customers were advised to complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recall@teleflex.com. Customers with no affected stock should also complete the enclosed Recall Acknowledgment Form to confirm receipt of the letter. Customers with questions should contact Customer Service at 1-866-246-6990.
Quantity
102,843 units
Distribution
Nationwide Distribution to CA, CO, FL, GA, IL, KS, KY, MA, MO, MS, NC, NJ, NY, OH, OK, OR, PA and TX.
Lot, serial or model codes
Product Code: 10096080, Lot number: 74F1501853 (8 FR); Product Code: 10096120, Lot numbers: 74E1602795, 74F1503201, 74G1600723 & 74J1500576 (12 FR); Product Code: 10096140, Lot numbers: 74A1600023, 74A1601105, 74A1602234, 74D1602007, 74E1602235, 74E1602796, 74F1501855, 74G1600486, 74H1500530 & 74L1500363 (14 FR); Product Code: 10096100, Lot numbers: 74A1600799, 74A1600800, 74A1602236, 74A1603122, 74D1600231, 74D1600923, 74D1601540, 74D1602136, 74E1600845, 74E1601768, 74E1602237, 74F1501857, 74F1501962, 74F1601966, 74F1601988, 74H1500534, 74H1500978, 74J1501157, 74K1502306, 74L1500364, 74L1501303, 74L1501938, 74M1500227 and 74M1501087 (10 FR) and Product Code: 10096160, Lot numbers: 74A1600024, 4A1600801, 74A1601106, 74A1603123, 74D1601541, 74E1600846, 74E1602797, 74G1600724, 74H1500979, 74J1501158, 74K1501837, 74L1501202 &74L1501939 (16 FR).
510(k) numbers
["K010420"]
PMA numbers
not on file
Product code
KOD
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register