Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer M/L Taper Rasp Handle, Right, 45 degrees, non-sterile; Part number 83-7712-035-02. Recall Z-0671-06 extended to include this product.

Recall number
Z-0771-06
Recalling firm
Zimmer Inc.
Product
Zimmer M/L Taper Rasp Handle, Right, 45 degrees, non-sterile; Part number 83-7712-035-02. Recall Z-0671-06 extended to include this product.
Status
Terminated
Initiated
2006-02-02
Posted
2006-03-21
Terminated
2007-12-05
Reason
A c-clip located near the rasp connection end of the instrument may detach during use and fall into the patient.
Root cause
Other
Action
Firm sales personnel were notified by letter dated 2/8/06 to visit each of their affected accounts and to remove the c-clip from the rasps or to remove the instrument from the hospital and return it to the firm. Similar instructions issued by letter dated 2/27/06 when the recall was extended to the two additional products.
Quantity
32 right and Left
Distribution
Nationwide, Australia, Canada, Germany, Korea, Singapore and Taiwan.
Lot, serial or model codes
Lots 08686600 and 08686800.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register