FDA Medical Device Recalls (openFDA)
GE Medical Systems, LLC: GE Healthcare, Brivo XR285amx Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
- Recall number
- Z-0769-2016
- Recalling firm
- GE Medical Systems, LLC
- Product
- GE Healthcare, Brivo XR285amx Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
- Status
- Terminated
- Initiated
- 2015-12-23
- Posted
- 2016-02-08
- Terminated
- 2016-09-07
- Reason
- Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
- Root cause
- Device Design
- Action
- Consignees were sent on 12/23/2015 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 10906 dated December 18, 2015. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers may continue to use the system. The firm will correct the affected devices and no charge to the customer; and a GE Healthcare representative will contact customers to arrange for the correction. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 126
- Distribution
- Worldwide Distribution -- US, including Washington D.C. Puerto Rico, Virgin Islands, and Guam; and the countries of Afghanistan, Albania, Anguilla, Antigua and Barbuda, Armenia, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, Ethiopia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Lebanon, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Papua New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Trinidad , Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
- Lot, serial or model codes
- Lot or serial numbers include but not limited to: 00000086177HL3, 00001028564WK1, 115226HL3, 00000083538HL9, 00000092940HL6, 00000084909HL1, 00000081798HL1, 120634HL1, 00000084729HL3, 00000084161HL9, 00000082463HL1, 00000080868HL3, 120832HL1, 00000089868HL4, 00000082736HL0, 00000093379HL6, 00000092616HL2, 00000092633HL7, 00000081160HL4, 00000077253HL3, 00000076913HL3, 00000078498HL3, 00000084160HL1, 00000078315HL9, 00000082726HL1, 00000085039HL6, 00000080635HL6, 00000091403HL6, 00000085598HL1, 00000085661HL7, 00000077923HL1, 00000081481HL4, 00001027046WK0, 00000081152HL1, 00000086364HL7, 00000092622HL0, 00000085000HL8, 00000084999HL2, 00000084725HL1, 00000083029HL9, 00000078313HL4, 00000084162HL7, 00000083020HL8, 00000077471HL1, 00000083019HL0, 00000082738HL6, 00000083598HL3, 00000092618HL8, 00000084180HL9, 00000085594HL0, 00000084917HL4, 00000076790HL5, 00000091413HL5, 00001028931WK2, 00000102856WK4, 00001028927WK0, 00001028924WK7, 00000084159HL3, 00000084422HL5, 00001029525WK1, 00000077482HL8, 00000082250HL2, 00000075962HL1, 00000079632HL6, 00000082240HL3, 00000081484HL8, 00000084732HL7, 00000077924HL9, 00000079918HL9, 00000082247HL8, 00000082245HL2, 00000082243HL7, 00000077255HL8, 00000086373HL8, 00000087093HL1, 00000102539HL4, 00000085004HL0, 00000085599HL9, 00000079559HL1, 00000079282HL0, 00000077782HL1, 00000082480HL5, 00000079116HL0, 00001028894WK2, 00000078317HL5, 00000079280HL4, 00000129740WG7, 00000079278HL8, 00000078492HL6, 00000075947HL2, 00000084908KL3, 00000084412HL6, 00000084169HL2, 00001027692WK1, 00000085036HL2, 0000000012C559, 00000084178HL3, 00000093852HL2, 00000092058HL7, 00000087079HL0, 00001029020WK3, 00001028888WK4, 00001028659WK9, 00001028662WK3, 00001028661WK5, 00000086380HL3, 00001028660WK7, 00000084172HL6, 00001028595WK5, 00000086365HL4, 00000084411HL8, 00000084402HL7, 00000084906HL7, 00001028866WK0, 00001028925WK4, and 00000085034HL7.
- 510(k) numbers
- ["K103476"]
- PMA numbers
- not on file
- Product code
- IZL
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28