Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: MiniCap Extended Life PD Transfer Set

Recall number
Z-0768-2021
Recalling firm
Baxter Healthcare Corporation
Product
MiniCap Extended Life PD Transfer Set
Status
Terminated
Initiated
2020-11-10
Posted
not on file
Terminated
2023-03-24
Reason
Potential for no-flow and leaks under the twist clamp.
Root cause
Process change control
Action
Urgent Medical Device Recall FA-2020-055 was sent to Peritoneal Dialysis Centers via U.S.P.S., first class mail on November 10, 2020. Urgent Medical Device Recall FA-2020-055 was sent via U.S.P.S., first class mail on November 13, 2020 to Home Patients who received impacted lots directly from Baxter. Out of an abundance of caution, Baxter will also be mailing the communication to all home patients who receive treatment at Peritoneal Dialysis Centers which received the impacted lots of transfer sets, regardless if the patient received the set or not.
Quantity
72,771 units
Distribution
Worldwide and US (Nationwide), Canada, Brazil, Mexico, Bermuda, and Indonesia
Lot, serial or model codes
Product Code: 5C4482, UDI: 00085412007731, Lot#, Expiration Date: H20D13100, 4/13/2025; H20D27084, 4/27/2025; H20D30070, 4/30/2025; H20E13107, 5/13/2025; H20E19054, 5/19/2025; H20E22082, 5/22/2025; H20F09061, 6/9/2025; H20F15050, 6/15/2025; H20F18070, 6/18/2025; H20F24086, 6/24/2025; H20F29051, 6/29/2025; H20G13046, 7/13/2025; H20G27087, 7/27/2025; H20G30065, 7/30/2025; H20H05057, 8/5/2025; H20H14067, 8/14/2025; H20H19066, 8/19/2025
510(k) numbers
["K192705"]
PMA numbers
not on file
Product code
KDJ
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28